TT-210 / Instrument QC source log Instrument ID: UTM-910 Serial number: S/N-7714 Check date: 2026-10-30 Technician: A. Patel Reference standard: STEP-210-A No QC sign-off or release has been issued. Check Id: QC-210-05 Reference Mm: 15.0 Tolerance Mm: 0.05 Reading Time: 2026-10-30 07:35 Evidence Ref: QC-LOG-05 Remarks: Observation withheld pending recheck Check Id: QC-210-03 Reference Mm: 10.0 Observed Mm: 9.99 Tolerance Mm: 0.05 Reading Time: 2026-10-30 07:21 Evidence Ref: QC-LOG-03 Remarks: Check standard log Check Id: QC-210-01 Reference Mm: 5.0 Observed Mm: 5.01 Tolerance Mm: 0.05 Reading Time: 2026-10-30 07:07 Evidence Ref: QC-LOG-01 Remarks: Check standard log Check Id: QC-210-04 Reference Mm: 12.0 Observed Mm: 12.07 Tolerance Mm: 0.05 Reading Time: 2026-10-30 07:28 Evidence Ref: QC-LOG-04 Remarks: Check standard log Check Id: QC-210-02 Reference Mm: 8.0 Observed Mm: 8.03 Tolerance Mm: 0.05 Reading Time: 2026-10-30 07:14 Evidence Ref: QC-LOG-02 Remarks: Check standard log Check Id: QC-210-06 Reference Mm: 20.0 Observed Mm: 19.98 Tolerance Mm: 0.05 Reading Time: 2026-10-30 07:42 Evidence Ref: QC-LOG-06 Remarks: Check standard log